Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Biochemistry Analyzer |
Still deciding? Get samples of US$ 6300/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications:
--Voltage of power source: AC 220V±22V, 50Hz±1Hz;
--UV wavelength: 300nm (transmission), 254 nm (repercussion), 365 nm (repercussion);
--Filter size (UV light): transmission window: 200mm x 150mm, Reflex window: 180mm x 40mm; Filter size (white light): 230mm x 190mm;
--SNR: < 66 dB;
-- Sensitivity: > 1 pg EB tinct double strand DNA;
-- Resolving capability:710 tenthousand pixel (3072pix ×2304pix ); Density of the picture: 24 Bit 16777216 grayscale;
--Continuous exposure time: 0.0005 ~ 900s;
--Have the functions of the softwares of 1D,2D,RFLP at the same time;
--The digital camera can take the pictures of the gel and adjust the pictures (Real-time preview ) via being connected with the computer, the Analysis Software can edit the images, identify the lanes automatically, calculate molecular weight and the distribution of the sample,etc.;
--Size (L X W X H):490×360×430 (mm);
-- Weight: about 21 kg
Applications:
WD-9413A is used for analyzing and researching into the gel image after the electrophoresis experiment. You can use the digital camera to take the picture of the gel under the UV light or the white light and then input the picture into the computer. With the help of the relevant special analysis software, you can analyse the images of DNA, RNA, protein gel, thin-layer chromatography etc. once and for all. And finally you can get the peak value of the band, molecular weight or base pair, area, height, position, volume or the total number of the samples.