Introduction
Blood gas analyzer can provide a safer alternative to follow ACCP [4] suggested "pH is chemical process optimization of pleural effusion, and pH must be conducted using a blood gas analyzer determination".BG 800 wise blood gas analyzer has a pH measurement of pleural effusion dedicated mode, can provide accurate measurement results, at the same time reduce the wrong blood pH pH mistaken for pleural effusion of risk.
Feature
Multi-combination parameters
Self-made maintenance free electrodes
Both for syringes & capillary
35UL for capillary, 95UL for whole blood
1 min from the aspiration to result
Long-lasting fluid system
All-in-one design calibrate cartridge
Cartridge expiration alarm & residue
Preheats reagents & sample
Optional Auto QC cartridge with H/M/L level
Data storage >5,000
Colorful touch screen with high resolution
Friendly operation interface
Infrared human-detector
Rich data management interface
Battery backup
Build-in thermal printer
Convenient bar-code system management
Economic cost per test
Technical Parameters
Analyte |
Measuring range |
Unit |
pH(CH) |
6.000-9.000 |
|
pCO2 |
1.607-26.67 |
kpa |
8.0-200.0 |
mmHg |
|
pO2 |
0-106.7 |
kpa |
0-800.0 |
mmHg |
|
K |
0.5-15.0 |
mmol/L |
Na+ |
20.0-200.0 |
mmol/L |
CI- |
20.0-200.0 |
mmol/L |
Ca++ |
0.3-5.0 |
mmol/L |
Hct |
12.0-65.0 |
% |
Item:
|
Blood & Gas Analysis System
|
Type:
|
Blood & Gas Analysis System
|
Certificate:
|
CE ISO13485
|
OEM & ODM:
|
Acceptable
|
Test Menu:
|
pH, pCO2, pO2, K, Na, Cl, Ca, Hct
|
Sample:
|
Whole blood, Serum, Plasma, Dialysate, CSF
|
Input Parameter:
|
Patient Temperature, Hemoglobin, FIO2, B.P.
|
Interface:
|
1. Serial Line RS-232.
2. TCP/IP interface, LIS connection under HL7 protocol.
3. USB port.
4. Integrated Bar-code reader.
|
Analysis Time:
|
<90s
|
Sample Volume:
|
1. 95uL whole blood (120uL if with Glu and Lac).
2. 50uL capillary blood.
|
Calibration:
|
1. Automatic.
2. Programmable 1-point and 2-point calibration.
|
Display:
|
10.4'' TFT touch screen
|
Battery:
|
Nickel-metal Hydride Battery, 3.0Ah
|
Printer:
|
Built-in Thermal Printer, 70mm
|
Barcode Reader:
|
Up to 14 Kinds Barcode Type
|
Work Environment:
|
15ºC~30ºC
|
Storage Environment:
|
5ºC~40ºC
|
Power Supply:
|
100-240V 50/60HZ
|
Instrument Dimension:
|
400(W)*574(H)*344(D) mm3
|
CAL PACK:
|
258(W)*180(H)*70(D) mm3
|
AutoQC Cartridge:
|
103(W)*180(H)*70(D) mm3
|
Instrument Weight:
|
18kg
|